Resistance meter calibration tih dan tur .

 May 20, 2023

 Message pakhat dah la .

 

1. Range chu 10 3 atanga 10 12 impedance bridge dik tak 1% a ni. A dikna sang Relative humidity meter (relative humidity hygrometer) thermometer dik tak (a dikna sang tak thermometer)
2. Cover chu hawng la, circuit board-a power switch-a wire pahnih inzawmte chu tihchhiat loh nan fimkhur rawh.
3. Circuit board hnuai dinglama calibration regulator pathum (calibration pots) zawng zawng zawng rawh .
4. He environmental condition hnuaiah hian darkar 1/2 tal watch chu dah la, self-balancing hnuah test tan rawh.
5. Tester nena rawn kal cable hmang la, a tawp khat chu alligator clip-ah leh a tawp dang chu banana plug-ah connect rawh.
6. 3.5 mm a sei plug chu meter socket-ah dah la.
7. Resistor tawp ber pahnih chu alligator clip hmangin connect rawh.
8. Calibration regulator pathum, a chung ber chu "humidity" tehna a ni a, a lai ber chu "impedance" a ni a, a hnuai ber chu "temperature adjustment with a small screwdriver" a ni. Clockwise direction chu value adjustment tihpunna atan a ni a, counterclockwise direction chu value adjustment tihtlem nan a ni.
9. Power switch chu hmet la, "temperature", "humidity" leh "resistance" value te chu a rualin khaikhin rawh.
10. Power switch chu thlah la, a inmil tur correction regulator chu zawi zawiin siam rawh.
11. Power switch chu hmet leh la, LCD display chu en rawh.
12. Recalibration a ngai a nih chuan power switch chu press la, calibrator chu siam leh rawh.
13. A khuhna chu dah la, fixing screw pali chu khauh takin khih rawh.
14. Power switch chu hmet la, watch chu a thawk tha em tih finfiah rawh.

Inquiry thawn rawh .
Inquiry thawn rawh .