Very Low Frequency (VLF) Cable Tester-a fault chinfel dan .

 Jun 20, 2025

 Message pakhat dah la .

 

Ultra-low-frequency generator te tan fault diagnosis method hrang hrang a awm a, chung zingah chuan power supply issues, abnormal high-voltage systems, protection functions triggering, leh test sample-a short circuit te pawh a tel. Common fault thenkhat leh a inmil diagnosis methods te hi han chhui ila.

 

1. Power on hnuah digital display a eng lo.
Power connection leh fuse a thawk tha em tih enfiah rawh. Fuse a chhiat chuan specification ang chiah fuse thar thlak a ngai a ni.

 

2. High-voltage indicator light a eng lo va, high voltage pawh a apply theih loh.
Zero Return Issue: Zero position a awm loh chuan high-voltage output adjustment knob leh fine-tuning knob chu zero position-a kir leh turin counterclockwise-in han herh leh teh. Tichuan, high voltage hman tum theih a ni.
Ground Wire Issue: Ground wire chu a inzawm dik em tih enfiah tur a ni.
Cable Connection Issue: Control box leh high-voltage tower inkar cable chu nghet taka inzawm a nih leh nih loh enfiah leh nemnghet rawh. Looseness emaw connection dik lo emaw a awm chuan cable chu connect leh la, hmanrua chu restart leh rawh.
Soldering Break Issue: Control box-a wire terminal pakhat chu solder leh disconnect a nih chuan hmanrua a restart hmain a resolder a ngai a ni.

 

3. Rated voltage a thlen hmain protection a trigger hmasa thin.

Overcurrent issue: Hei hi short circuit emaw leakage current tam lutuk vang emaw a ni thei a, chu chuan large-capacity loads a hmachhawn hunah voltage a tisang rang hle. A solution chu hmanrua restart leh voltage rise speed tih hniam a ni.
Overvoltage Issue: Overvoltage protection setting knob chu a tam berah a awm loh chuan, a awmna hmun sang berah darkar lam hawiin rotate la, chutah chuan hmanrua chu restart leh rawh.

 

4. Test object-ah short circuit a awm a, voltage sang tak hnuaiah a awm .
Test object hian voltage sang tak hnuaia short circuit a tawn chuan short tam tak-circuit current avang hian fault indicator light zawng zawng a eng vek thei a ni. Hei hi thil thleng pangngai a ni. Test object troubleshoot hnuah hmanrua chu restart leh la, normal operation chu i tan leh thei ang.

 

5. Digital reading hian inthlak danglamna lian tak a nei a.
Voltmeter digital reading a inthlak danglam nasa hle a nih chuan a tawp pahniha shielding wire head leh connecting cable central shielding wire head te chu a inzawm lo em tih enfiah tur a ni. Harsatna eng emaw a awm chuan welding hmangin rang takin siam tha leh rawh.

 

Inquiry thawn rawh .
Inquiry thawn rawh .