DC Hipot Tester Calibrate dan tur .

 Mar 17, 2025

 Message pakhat dah la .

 

DC hipot test equipment calibration hi test result dik leh rintlak a nih leh nih loh enfiahna atana thil pawimawh tak a ni. A hnuaia tarlan te hi calibration atana ruahmanna leh fimkhur dan chipchiar tak chu a ni:

1. Calibration hmaa inbuatsaihna .
Device enfiah rawh: Device chu a dinhmun a tha em tih enfiah la, chhiatna leh tarna langsar tak tak awm lovin.
Environment dinhmun: Temperature leh humidity dik tak awmna hmunah calibration tih tur a ni a, a tlangpuiin 23±5℃leh humidity 75% aia tlem emaw, a tlukpui emaw a ni tur a ni.
Standard Equipment: High-Precision Standard Voltmeter, Standard Ammeter, emaw Professional Calibrator emaw chu test mek hmanrua aiin a dikna a sang zawk em tih enfiah tur a ni.

2. Thil chi hrang hrang calibrate .
DC hipot test hmanruaa calibration item lian ber berte chu:

Output voltage calibration: Device hmanga DC voltage output chu a dik em tih finfiah rawh.
Leakage Current Measurement Calibration: Tuna tehna hnathawh dik leh dik loh finfiah.
Time Control Calibration: Test time control dik leh dik loh finfiah.

3. Calibration tih dan tur .
(a) Output voltage calibration 1.1.
Standard voltmeter chu device output ah connect rawh.
Device output voltage chu value hrang hrang (1KV, 5KV, 10KV, etc. ang chi) ah set la, standard voltmeter chhiarna chu record rawh.
Device display value chu standard value nen khaikhin la, error chu chhut la, error chu allowable range chhungah a awm em tih enfiah rawh (a tlangpuiin ±1% aia tlem emaw a tlukpui emaw a ni).
(b) Leakage current tehna calibration .
Standard current source emaw high-precision ammeter chu device output end ah connect rawh.
Device-a output voltage fixed chu set la, device-in a rawn tarlan leakage current value leh standard value chu record rawh.
An pahnih hi tehkhin la, error chu allowable range chhungah (a tlangpuiin ±2% aia tlem emaw, a tlukpui emaw) a awm theih nan.
(c) Hun hman dan (time control) calibration .
Equipment-a time control function test nan high-precision timer emaw calibrator emaw hmang la.
Test time hrang hrang (minute 1 leh minute 5 ang chi) set la, a hun tak tak leh device-a hun tarlan chu record rawh.
An pahnih hi tehkhin la, error chu allowable range chhunga awm (a tlangpuiin ±1 second aia tlem emaw, a tlukpui emaw) a nih leh nih loh enfiah rawh.

4. Calibration record leh report .
Record data: Calibration process-a input value, output value, error leh data dangte chu chipchiar takin record rawh.
Report siam: Calibration result hmanga calibration report siam, hmanrua chungchanga hriat tur, calibration ni, calibration result, error analysis, etc. te pawh a tel.
Identification status: Thil hman thiamte chunga calibration hriat theihna paste, calibration hman hun chhung a tarlang.

DC hipot test equipment calibration hi test dik leh hmanrua rintlakna tur atana link pawimawh tak a ni. Standardized calibration process leh strict error control hmangin, test tak takah chuan hmanrua te chu a tha ber tih theih a ni a, chutih rualin a kaihhnawih standard leh specification-a thil tulte chu a tlin theih bawk.

Inquiry thawn rawh .
Inquiry thawn rawh .